A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software.                                                                                                                                                                                                Other related CVE's are CVE-2025-14095 & CVE-2025-14096.                                                                                                      Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer.                                                                                                                                                                        Temporary work Around: If the network is not considered secure, please remove the analyzer from the network.                         Permanent solution: Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution.                                                      Exploit Status: Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.
References
History

Fri, 19 Dec 2025 00:15:00 +0000

Type Values Removed Values Added
Metrics ssvc

{'options': {'Automatable': 'no', 'Exploitation': 'none', 'Technical Impact': 'total'}, 'version': '2.0.3'}


Thu, 18 Dec 2025 10:00:00 +0000

Type Values Removed Values Added
First Time appeared Radiometer
Radiometer abl800 Basic Analyzer
Radiometer abl800 Flex Analyzer
Radiometer abl90 Flex Analyzer
Radiometer abl90 Flex Plus Analyzer
Radiometer aqt90 Flex Analyzer
Vendors & Products Radiometer
Radiometer abl800 Basic Analyzer
Radiometer abl800 Flex Analyzer
Radiometer abl90 Flex Analyzer
Radiometer abl90 Flex Plus Analyzer
Radiometer aqt90 Flex Analyzer

Wed, 17 Dec 2025 13:00:00 +0000

Type Values Removed Values Added
Description A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software.                                                                                                                                                                                                Other related CVE's are CVE-2025-14095 & CVE-2025-14096.                                                                                                      Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer.                                                                                                                                                                        Temporary work Around: If the network is not considered secure, please remove the analyzer from the network.                         Permanent solution: Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution.                                                      Exploit Status: Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.
Title Remote Code Execution Vulnerability in Radiometer Products
Weaknesses CWE-287
References
Metrics cvssV3_1

{'score': 7.2, 'vector': 'CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H'}


cve-icon MITRE

Status: PUBLISHED

Assigner: Radiometer

Published:

Updated: 2025-12-17T14:41:13.252Z

Reserved: 2025-12-05T10:50:03.683Z

Link: CVE-2025-14097

cve-icon Vulnrichment

Updated: 2025-12-17T14:41:04.401Z

cve-icon NVD

Status : Awaiting Analysis

Published: 2025-12-17T13:15:58.050

Modified: 2025-12-18T15:08:06.237

Link: CVE-2025-14097

cve-icon Redhat

No data.

cve-icon OpenCVE Enrichment

Updated: 2025-12-18T09:57:32Z